--- author: Sam White toc-title: Contents toc-depth: 5 toc-location: left date: 2015-11-25 02:03:04+00:00 layout: post slug: phenol-chloroform-dna-cleanup-geoduck-gdna title: Phenol-Chloroform DNA Cleanup - Geoduck gDNA categories: - 2015 - Geoduck Genome Sequencing tags: - DNA Quantification - gDNA - geoduck - NanoDrop1000 - Panopea generosa - phenol-chloroform cleanup - Quant-iT DNA BR Kit - Victor 1420 plate reader --- The [gDNA I extracted on 20151104](https://robertslab.github.io/sams-notebook/posts/2015/2015-11-05-dna-quantification-quality-assessment-geoduck-oly-gdna/) didn't look great on the NanoDrop so I decided to perform a phenol-chloroform cleanup to see if I could improve the NanoDrop1000 absorbance spectrum and, in turn, the quality of the gDNA. * Added an equal volume (500μL) of phenol:chloroform:isoamyl alcohol (25:24:1) to the DNA * Mixed by hand - moderate shaking * Centrifuged 2mins, 10,000g, RT * Transferred aqueous phase to clean tube and discarded interphase & organic phase * Added an equal volume (280μL) of chlforoform:isoamyl alcohol (24:1) * Mixed by hand - moderate shaking * Centrifuged 2mins, 10,000g, RT * Transferred aqueous phase (210μL) to clean tube * Added 0.1vols (21μL) of 3M sodium acetate (pH = 5.2) * Added 2vols (420μL) of 100% EtOH * Mixed by inversion * Incubated @ -20C, 1hr (probably not necessary since gDNA clearly precipitated out as soon as I mixed the sample) * Pelleted DNA by centrifuging 15mins, 12,000g, RT * Discarded supe * Washed pellet with 1000μL cold (-20C) 70% EtOH * Centrifuged 5mins, 12,000g, RT * Discarded supe * Repeated was steps three more times * Resuspended pellet in 100μL of Buffer EB (Qiagen) DNA was quantified using two methods: NanoDrop1000 & QuantIT dsDNA BR Kit For the Quant-IT kit, the samples were quantified using the QuantIT dsDNA BR Kit (Invitrogen) according to the manufacturer's protocol. Standards were run in triplicate, samples were run in duplicate. 96-well black (opaque) plate was used. Fluorescence was measured on the Seeb Lab's Victor 1420 plate reader (Perkin Elmer). Results:
METHOD CONCENTRATION (ng/μL) VOLUME (μL) YIELD (ng)
NanoDrop1000 371.83 100 37,183
Quant-IT 100.83 100 10,082
The NanoDrop1000 overestimates the concentration of the sample by 3.7x! Regardless, the yield isn't all that great (using yield from Quant-IT), which has generally been the case for all of the geoduck gDNA isolations I've performed. It would probably be prudent to try isolating gDNA from a different tissue to see if yields improve... Will evaluate gDNA quality on a gel. Fluorescence (Google Sheet): [20151124_geoduck_oly_gDNA_quants](https://docs.google.com/spreadsheets/d/167If9r5fDNJb6xenUo-bKy52794_O1LgN1twKAOmElw/edit?usp=sharing) NanoDrop1000 Measurements and Plots The clean up procedure didn't really seem to help with the geoduck sample, as we're still seeing a significant amount of absorbance from 230 - 250nm. ![](https://eagle.fish.washington.edu/Arabidopsis/20151124_gDNA_geoduck_oly_ODs.JPG)(http://eagle.fish.washington.edu/Arabidopsis/20151124_gDNA_geoduck_oly_ODs.JPG) ![](https://eagle.fish.washington.edu/Arabidopsis/20151124_gDNA_geoduck_oly_plots.JPG)(http://eagle.fish.washington.edu/Arabidopsis/20151124_gDNA_geoduck_oly_plots.JPG)