--- author: Sam White toc-title: Contents toc-depth: 5 toc-location: left date: 2009-08-24 16:10:32+00:00 layout: post slug: qpcr-additional-calibration-test-of-opticon-2 title: qPCR - Additional Calibration test of Opticon 2 categories: - 2009 - Miscellaneous tags: - calibration - Opticon2 - qPCR - troubleshooting --- Based on recs from Bio-Rad rep (Carl Fisher), will repeat Opticon 2 calibration (see [20090813](/Sam%27s+Working+Notebook+Jun-Aug+2009#sjw20090813)) test according to Opticon manual. Then, will rotate plate 180 degrees and repeat test and upload data to Bio-Rad server for analysis and evaluation. Results: According to the Bio-Rad rep, he thinks recalibration is in order. He pointed out that we're seeing a 3-fold spread in fluorescence, which is outside of the "tolerable" range (which is 2-fold). Will recalibrate.